The Geoform disease-specific annuloplasty system: a finite element study
Emiliano Votta1, Francesco Maisano, Steven F Bolling
1Politecnico di Milano, Bioengineering Department, Cardiac Surgery Division, IRCCS San Raffaele Hospital, Milan, Italy. emiliano.votta@polimi.it
A new Geoform ring for functional mitral regurgitation (FMR) showed superior results compared to standard annuloplasty rings. This FMR-specific device improves leaflet coaptation and reduces stress, offering more durable outcomes for patients.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Medical Device Design
Background:
- Functional mitral regurgitation (FMR) involves impaired mitral leaflet coaptation due to functional and geometric issues.
- Standard annuloplasty, while common, may be insufficient for severe FMR.
- FMR-specific rings offer a potential for improved surgical outcomes.
Purpose of the Study:
- To compare the performance of the novel Geoform ring with a standard Physio annuloplasty ring for FMR correction.
- To evaluate the efficacy of these devices in simulated FMR conditions using finite element modeling.
Main Methods:
- Finite element modeling simulated dilated cardiomyopathy-related FMR with varying degrees of papillary muscle displacement.
- The Geoform and Physio annuloplasty rings were modeled in three sizes to assess their impact on valve dynamics.
- Key parameters evaluated included regurgitant area, coaptation length, and leaflet stresses.
Main Results:
- The Geoform ring consistently achieved coaptation lengths over 7 mm, with a 74% reduction in regurgitant area in severe cases.
- Leaflet stresses were reduced by 20% with the Geoform ring, even with simulated undersizing.
- The standard Physio ring resulted in a maximum coaptation length of 4.3 mm and a 60% regurgitant area reduction.
Conclusions:
- Disease-specific prostheses like the Geoform ring can effectively restore mitral valve competence in cases of significant leaflet tethering.
- These specialized devices may eliminate the need for aggressive undersizing, potentially leading to more durable surgical results.
- The Geoform ring demonstrates promising performance for addressing severe functional mitral regurgitation.