Related Experiment Video
Updated: Jul 14, 2026

The Intra-Aortic Balloon Pump
Published on: February 5, 2021
Transbrachial intraaortic balloon pumping in severe peripheral atherosclerosis
Francesco Onorati1, Barbara Impiombato, Alessandro Ferraro
1Cardiac Surgery Unit, Magna Graecia University, Catanzaro, Italy. frankono@libero.it
Abstract:
Preoperative intraaortic balloon pumping improves the results of complex coronary surgery; however, insertion may be harmful or contraindicated in severe and diffuse atherosclerosis of the descending aorta and peripheral arteries. We report our experience with 10 consecutive patients with severe peripheral atherosclerosis or distal abdominal aortic aneurysms, in whom a 7.5F intraaortic balloon catheter was inserted through the brachial artery. Intraaortic balloon pumping was maintained until hemodynamic stability was established; no complications or ischemia of the hand related to the intraaortic balloon pump occurred. Transbrachial intraaortic balloon pumping with a 7.5F catheter is as safe and effective as the transfemoral method in patients with unavailable femoral arteries.
More Related Videos
Related Concept Videos
Peripheral Artery Disease III: Interprofessional Care
Aortic Regurgitation III: Medical Management