Closed-loop anaesthesia delivery system (CLADS) using bispectral index: a performance assessment study.

G D Puri1, B Kumar, J Aveek

  • 1Department of Anaesthesiology and Intensive Care, Post Graduate Institute of Medical Education and Research, Chandigarh, India.

Anaesthesia and Intensive Care
|June 27, 2007
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Prize Winning Abstracts from BASEM Congress 2016.

British journal of sports medicine·2017
Same author

Nocardia Mycetoma At Unusual Site.

Indian journal of dermatology, venereology and leprology·2017
Same author

Tinea Cruris in Chandigarh.

Indian journal of dermatology, venereology and leprology·2017
Same author

Penicillin Resistant Gonococci.

Indian journal of dermatology, venereology and leprology·2017
Same author

Erythema Elevatum Diutinum with a Prolonged Bullous Phase.

Indian journal of dermatology, venereology and leprology·2017
Same author

Reiter's Disease- Clinical Profile of Six Cases.

Indian journal of dermatology, venereology and leprology·2017

Automated anesthesia delivery using a closed-loop system (CLADS) effectively controlled propofol and bispectral index (BIS) levels, leading to improved patient recovery compared to manual control.

Area of Science:

  • Anesthesiology
  • Medical Devices
  • Automation

Background:

  • Electroencephalographic (EEG) indices like the bispectral index (BIS) are crucial for monitoring anesthetic depth.
  • This has spurred interest in automated anesthesia delivery systems that utilize EEG data.

Purpose of the Study:

  • To compare a patented closed-loop anesthesia delivery system (CLADS) with manual propofol titration to BIS.
  • To evaluate the efficacy and efficiency of automated anesthesia delivery.

Main Methods:

  • Forty ASA I-II patients undergoing elective surgery were randomized into two groups.
  • One group received propofol via CLADS; the control group received manual propofol titration to BIS.

Main Results:

Related Experiment Videos

  • The CLADS group showed lower propofol induction doses and less BIS overshoot (P < 0.05).
  • BIS was maintained within +/-10 of target for longer periods in the CLADS group (P < 0.01).
  • The CLADS system required less anesthetic agent (P < 0.01) and resulted in faster postoperative recovery (P < 0.05).
  • Conclusions:

    • Automated propofol delivery using the CLADS is effective and efficient compared to manual titration.
    • The CLADS system reduces anesthesiologist workload by minimizing infusion rate adjustments.