Time-limited fluoroscopy to reduce radiation exposure during ERCP: a prospective randomized trial
Lance T Uradomo1, Eric M Goldberg, Peter E Darwin
1Division of Gastroenterology and Hepatology, University of Maryland Medical Center, Baltimore, Maryland 21201, USA.
Background:
Measures should be used to limit radiation exposure of the staff and patients during ERCP.
Objective:
To determine whether "time-limited fluoroscopy" reduces radiation exposure and fluoroscopy time (FT) compared with continuous fluoroscopy.
Design:
Prospective randomized trial.
Setting:
Tertiary academic medical center.
Patients:
Consecutive adult patients presenting for ERCP.
Interventions:
Subjects were randomized into 2 fluoroscopy setting groups: (1) time-limited, where x-ray exposure is limited to 3 seconds each time the foot-operated switch is depressed; (2) continuous, where x-ray exposure continues for as long as the switch is depressed.
Main Outcomes Measurements:
FT, patient, and procedure-related data were recorded. Radiation dosimetry badges were used to estimate cumulative exposure.
Results:
Ninety-nine procedures were performed in the time-limited group and 100 by using continuous fluoroscopy. The mean FT for time-limited fluoroscopy was 284.4 seconds (95% confidence interval [CI] 247.1-321.6) and for continuous fluoroscopy was 314 seconds (95% CI 265.6-362.4; P=.34). Longer FT was associated with moderate or difficult cannulation (P=.008), lithotripsy (P<.001), stent placement (P=.007), sphincterotomy (P<.001), and longer overall procedure length (P<0.001). After controlling for confounding factors and interactions with a multiple linear regression model, time-limited fluoroscopy was associated with a 16.4% lower FT (P=.029). The average radiation dose was not amenable to multivariate analysis, and, therefore, no significant difference between groups was found.
Limitations:
Endoscopists were not blinded to the study group assignments.
Conclusions:
FT and radiation exposure are dependent upon numerous patient-, operator-, and procedure-related factors. This study found that, after controlling for the impact of confounding factors, time-limited fluoroscopy significantly decreases FTs.
More Related Videos
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Endoscopic Procedures V: ERCP
Patient...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...