Related Experiment Video
Updated: Jul 14, 2026

Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
The development of a model for in vitro testing of femoral stems in impaction bone grafting
J S Heal1, C Gozzard, S Gheduzzi
1Department of Orthopaedics, University of Bristol, Bristol, UK. jsheal@doctors.org.uk
Abstract:
Comprehension of the biomechanical behaviour of orthopaedic implants is essential. This paper describes the development of an in vitro model to investigate the behaviour of femoral implants in the revision setting. The development of a femoral model and a bone graft substitute is described. The properties of human, bovine, ovine morselized bone graft, and a graft substitute were compared. On measuring hoop strain after impaction bone grafting there was no significant difference between the ovine bone graft and graft substitute with the size 1 Exeter stem. The results suggest that this bone graft substitute is a viable alternative for in vitro testing. The authors recommend the use of the graft substitute and the femoral model to predict femoral stem biomechanics.