Related Experiment Video
Updated: Jul 13, 2026

In Situ Surface Temperature Measurement in a Conveyor Belt Furnace via Inline Infrared Thermography
Published on: May 30, 2020
Infrared microthermography of microfabricated devices
Robert Furstenberg1, C A Kendziora, Stanley V Stepnowski
1U.S. Naval Research Laboratory, Code 6365, Washington, DC 20375, USA. robert.furstenberg@gmail.com
Abstract:
We report a new experimental apparatus for infrared microthermography applicable to a wide class of samples including semitransparent ones and perforated devices. This setup is particularly well suited for the thermography of microfabricated devices. Traditionally, temperature calibration is performed using calibration hot plates, but this is not applicable to transmissive samples. In this work a custom designed miniature calibration oven in conjunction with spatial filtering is used to obtain accurate static and transient temperature maps of actively heated devices. The procedure does not require prior knowledge of the emissivity. Calibration and image processing algorithms are discussed and analyzed. We show that relatively inexpensive uncooled bolometer arrays can be a suitable detector choice in certain radiometric applications. As an example, we apply this method in the analysis of temperature profiles of an actively heated microfabricated preconcentrator device that incorporates a perforated membrane and is used in trace detection of illicit substances.