Background:
Open tracheostomy is commonly performed during head and neck surgery, and in critically ill patients. Diathermy-induced airway fire during tracheotomy is rare but may have grave implications. Recommendations to minimize this risk are not always practical. We hypothesized that flooding the surgical field with carbon dioxide is an effective technique in preventing fire.
Methods:
We cut through the trachea of two pigs using diathermy while ventilating with pure oxygen five times with, and five times without, simultaneous flushing of the surgical field with carbon dioxide at 10 L/min. To increase the amount of oxygen in the airway and the likelihood of fire,we deliberately deflated the endotracheal cuff to simulate cuff rupture.
Results:
Five times out of five, fire was induced when the diathermy cut through the tracheal wall with no carbon dioxide being used. Five times out of five, fire was not induced when carbon dioxide was used. The difference was significant (p < 0.008).
Conclusions:
Flooding the surgical site with carbon dioxide effectively prevents fire during open tracheostomy using diathermy.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.