Optimization of extraction chromatography separations of trace levels of actinides with ICP-MS detection
Dominic S Peterson1, Alexander A Plionis, Edward R Gonzales
1Chemistry Division, Los Alamos National Laboratory, Los Alamos, NM 87545, USA. DominicP@lanl.gov
Abstract:
The separation of trace level actinides has been evaluated on extraction chromatography columns. Detection of the actinides was achieved through the use of an inductively coupled plasma MS (ICP-MS). The columns that we tested were prepared from a commercial TRU resin. The separation of the actinides was optimized for several parameters including particle size, column length, packing pressure, and eluent flow rate. We also examined the possibility of reducing or eliminating oxalic acid in the eluents in order to improve the performance of the mass spectrometer. We were able to separate a mixture of five actinides ((232)Th,( 238)U,( 237)Np, (239)Pu,( 243)Am) in less than 4 min. This work has application to rapid bioassay as well as for automated separations of actinide materials.
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Ion-Exchange Chromatography
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Atomic Emission Spectroscopy: Lab