Evaluation of optimal echo time for 1H-spectroscopic imaging of brain tumors at 3 Tesla

Elke Hattingen1, Ulrich Pilatus, Kea Franz

  • 1Institute of Neuroradiology, Johann Wolfgang Goethe University of Frankfurt/Main, Frankfurt, Germany. elke.hattingen@kgu.de

Abstract

Insights

Short echo time (TE) MR spectroscopic imaging (MRSI) at 3 Tesla offers superior signal-to-noise ratio and improved fitting accuracy for brain tumor metabolites compared to long-TE MRSI.

Area of Science:

  • Neuroimaging
  • Magnetic Resonance Spectroscopy
  • Biomedical Engineering

Background:

  • Magnetic Resonance Spectroscopic Imaging (MRSI) is crucial for non-invasively assessing brain tissue composition.
  • Optimizing MRSI parameters like echo time (TE) is essential for enhancing diagnostic accuracy.
  • Previous studies have explored various TE values, but direct comparison at 3 Tesla for focal brain lesions is needed.

Purpose of the Study:

  • To compare the spectral quality of short-TE (30 msec) and long-TE (144 msec) MRSI at 3 Tesla.
  • To evaluate the performance of these TE values in normal brain and tumor tissues.
  • To determine the optimal TE for analyzing key brain metabolites.

Main Methods:

  • 32 patients with brain lesions underwent 3 Tesla MRSI using chemical-shift imaging (CSI) at TE = 30 msec and TE = 144 msec.
  • Tumor-relevant metabolites including trimethylamine (TMA), creatine compounds (tCr), and N-acetylated compounds (tNAA) were quantified.
  • LCModel software was used for spectral analysis, employing prior knowledge fitting.

Main Results:

  • Short-TE MRSI demonstrated up to double the signal-to-noise ratio (SNR) compared to long-TE MRSI.
  • Fitting error was reduced by up to 30% for TMA and tCr with short-TE.
  • Absolute metabolite concentrations were consistent across both TE values, indicating reliable quantification.

Conclusions:

  • Short-TE MRSI (30 msec) is recommended as the preferred method at 3 Tesla for evaluating focal brain lesions due to superior spectral quality.
  • Long-TE MRSI (144 msec) may be a viable alternative in specific cases with potential baseline instability.
  • Further investigation into metabolites detectable at very short TEs (< 30 msec) could yield additional diagnostic insights.