Related Experiment Video
Updated: Jul 13, 2026

An “All-laser” Endothelial Transplant
Published on: July 6, 2015
Long-term stability of the posterior cornea after laser in situ keratomileusis
Joseph B Ciolino1, Stephen S Khachikian, Michael J Cortese
1Albany Medical College, Albany, New York, USA.
Purpose:
To study long-term changes in posterior corneal elevation after laser in situ keratomileusis (LASIK) using Scheimpflug topography (Pentacam, Oculus, Inc.) in eyes 1 year after LASIK.
Setting:
Department of Ophthalmology, Albany Medical Center, and a private practice, Albany, New York, USA.
Methods:
One hundred two myopic eyes of 52 consecutive patients presenting for their 1-year follow-up were prospectively evaluated using the Pentacam to determine elevation changes to the posterior corneal surface between preoperative and 1-year postoperative measurements. Changes in posterior elevation were performed by comparing the best-fit sphere preoperatively and postoperatively with a fixed reference sphere determined by the central 9.0 mm preoperative cornea. Statistical and graphical analyses were performed.
Results:
One hundred two post-LASIK eyes (mean correction -4.33 diopters; mean ablation depth 68.70 microm; mean estimated residual bed thickness 327 microm) had a mean posterior displacement of -0.47 microm +/- 3.48 (SD) (range -10.0 to +7 microm). The mean follow-up period was 13.6 months (range 8.8 to 19.3 months).
Conclusions:
In this population, no patient had significant forward protrusion of the posterior corneal surface a mean of 14 months after LASIK. The posterior cornea in post-LASIK myopic eyes was very stable. Contrary to results in previous studies, progressive changes to the posterior corneal surface did not routinely occur after LASIK performed within established parameters.