High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Fourier transform capillary electrophoresis with laminar-flow gated pressure injection
Peter B Allen1, Byron R Doepker, Daniel T Chiu
1Department of Chemistry, University of Washington, Seattle, WA 98195, USA.
Abstract:
Fourier transform capillary electrophoresis (FTCE) was developed as a method to improve signal-to-noise ratio (S/N) and resolution in capillary electrophoresis (CE) separation. In FTCE, multiple simultaneous CE separations were performed in the same channel system and interrogated using a single-point detector. To illustrate experimentally the improvement offered by FTCE in S/N ratio and resolution, we carried out a modest number (five) of multiple injections and separations. We show even with this small number of separations, S/N increased by a factor of 2.9, and theoretical plate height improved by a factor of more than 30. We demonstrated this technique with laser-induced fluorescence detection, but a wide variety of detection methods are compatible with FTCE.
More Related Videos
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Electrophoresis: Overview
There...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
High-Performance Liquid Chromatography: Elution Process
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such as cells...
High-Performance Liquid Chromatography: Instrumentation