Home monitoring with a modified automatic sphygmomanometer to detect recurrent atrial fibrillation
Joseph Wiesel1, David J Wiesel, Frank C Messineo
1New York Hospital Queens, Flushing, New York 11355, USA.
Objectives:
Detecting asymptomatic atrial fibrillation would help identify patients who would benefit from anticoagulation. This study examined the application of a modified blood pressure monitor to screen for episodes of atrial fibrillation.
Methods:
A modified sphygmomanometer was designed to detect atrial fibrillation. The device has a sensitivity near 100% and a specificity of up to 91%. Therefore, this device can be expected to detect all episodes of atrial fibrillation. However, the lower specificity may result in false-positive readings that could prompt unnecessary clinic visits for electrocardiogram confirmation of the rhythm. Outpatients in sinus rhythm with a history of atrial fibrillation were given the device to monitor their pulse regularity once daily to detect atrial fibrillation. Patients with irregular readings were evaluated with an electrocardiogram.
Results:
Nineteen patients were monitored at home for a period ranging from 5 days to 5 months. Seven patients had recurrent atrial fibrillation identified by the monitor. Nine patients had no irregular readings for a mean of 82 +/- 40 days. Of 19 patients, 3 had false-positive irregular readings that were a result of sinus arrhythmia or ectopy.
Conclusions:
The device had an acceptably low false-positive rate allowing 16 of 19 patients to use it at home for long-term atrial fibrillation monitoring. This device may help prevent strokes by identifying patients with prolonged episodes of asymptomatic atrial fibrillation who are candidates for anticoagulation.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Holter Monitor: 24-Hour Monitoring
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)