Abstract:
A high-performance liquid chromatographic method with ultraviolet photometric detection for the determination of cotinine in human plasma was described. The use of a 30-cm reversed-phase column and of a mobile phase consisting of water-methanol-0.1 M sodium acetate-acetonitrile (72:21:5.6:1.4, v/v), pH 4.1, eliminated caffeine interference. A simplified solid-phase extraction procedure was also performed for plasma samples.