Transthoracic impedance changes as a tool to detect malpositioned tracheal tubes
Jo Kramer-Johansen1, Joar Eilevstjønn, Theresa Mariero Olasveengen
1Institute for Experimental Medical Research, Ullevål University Hospital, N-0407 Oslo, Norway. jo.kramer-johansen@medisin.uio.no
Background:
Undetected malpositioned or dislodged ventilation tubes during cardiac arrest have fatal consequences, and no single method can detect the tube position reliably during such low-flow states. We wanted to test the ability of impedance changes as measured across the chest via the standard defibrillation pads to distinguish between oesophageal and tracheal ventilations in non-circulated patients.
Materials And Methods:
After the end of futile resuscitation transthoracic impedance was measured with a prototype defibrillator, and ventilation variables were collected with a spirometer-capnography unit during tracheal ventilations and after repositioning of the tube; during oesophageal ventilations for paired comparisons.
Results:
We registered 123 oesophageal and 178 tracheal ventilations in nine patients. Transthoracic impedance changes associated with ventilations were always larger during tracheal than oesophageal ventilations (mean difference 1.3 ohms (95% CI 1.0, 1.5), P<0.001), and all such changes above 1.2 ohms were associated with tracheal ventilations, while changes below 0.4 ohms always were associated with oesophageal ventilations. By subtracting 0.5 ohms from the individual mean transthoracic change associated with tracheal ventilations, tube position was predicted with sensitivity 0.99 and specificity 0.97.
Conclusion:
Transthoracic impedance changes may be used to detect malpositioned and dislodged tubes also during situations without spontaneous circulation. Our predictive values must be retested in another population.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Physical Assessment of the Respiratory Tract II: Palpation
Thoracic Palpation
Thoracic palpation detects tenderness, masses, lesions, respiratory excursions, and vocal fremitus. The nurse assesses...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: