Related Experiment Video
Updated: Jul 12, 2026

In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
Intravenous fluid therapy: a randomized controlled trial to investigate the effectiveness of the IV(2) flow medical
Nesta Fraser1, Jacques R Snyman, Francois Wessels
1Department of Pharmacology, University of Pretoria, Pretoria, Gauteng, South Africa. nestal@absamail.co.za
Objectives:
To investigate the hypothetical benefits of the IV(2) flow medical device.
Background:
Intravenous fluid administration is a standard hospital procedure with assumed inadequacies. The IV-Event Study [Fraser N, Nel G, Snyman J & Wessels F (2004) IV-EVENT Study: Intravenous Infusion Therapy--Management and Adverse Events. Data on File: Varori International (Pty) Ltd., Centurion, South Africa] quantified these inadequacies; The 'Stargait' intervention trial investigated the effectiveness and possible cost-benefit of the IV(2) flow. The IV(2) flow is intended for routine use with gravitational intravenous infusion sets. The IV(2) flow should reduce the incidence rate of adverse events and maintain a set flow rate.
Method:
Nursing staff assisted by study assessors captured relevant data. Consented patients were enrolled for the period of their prescribed infusions. Intervention. The Stargait Trial compared the treatment group (standard gravitational sets with the IV(2) flow) with the control group (standard gravitational infusion sets without IV(2) flow). The difference in observed events and the cost benefit derived from this were measured.
Results:
A total of 2387 drip hours were observed in 52 patients. The adverse event rates were: Control group (without IV(2) flow) 33.8%. The treatment group (IV(2) flow) 15.4%. This 55% reduction is statistically significant (p = 0.0069). Adverse event related monetary wastage (labour and consumables) is reduced by 76% for infusion bags in the intervention group (with IV(2) flow). There was a significant difference between the treatment group and control group as far as deviation in flow rate was concerned (p = 0.00818). The mean deviation of the IV(2) flow group was just more than 5 ml per hour. The standard line group had a mean deviation of more than 30 ml per hour.
Conclusion:
Gravitational intravenous therapy compromises quality of patient care. The Stargait Trial has proven the care-effectiveness and cost-effectiveness of the IV(2) flow.
Relevance To Clinical Practice:
The IV(2) flow improves quality of patient care and reduces associated wastage.
More Related Videos
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Heart Failure VI: Adjunct Therapies