Hypersensitivity reactions to ethylene oxide: clinical experience
G Masin1, M Polenaković, N Ivanovski
1Department of Nephrology, Medical Faculty, Skopje, Yugoslavia.
Summary
Hypersensitivity reactions during hemodialysis were linked to ethylene oxide (ETO) in new dialyzers. Higher ETO levels in dialyzer potting suggest a potential cause for these acute patient responses.
Area of Science:
- Nephrology
- Clinical Chemistry
- Biomaterials Science
Background:
- Hypersensitivity reactions can occur during hemodialysis, particularly with new dialyzer use.
- Ethylene oxide (ETO) sterilization is common for medical devices, but residual ETO can pose risks.
Purpose of the Study:
- To investigate the cause of acute hypersensitivity reactions observed in hemodialysis patients.
- To determine the association between ETO concentration in dialyzers and the occurrence of these reactions.
Main Methods:
- Observational study of hemodialysis patients experiencing acute reactions.
- Analysis of hollow-fibre dialyzers (cellulose ester and cuprammonium cellulose) for ETO concentration.
- Correlation of ETO levels in dialyzer components (potting vs. capillaries) with clinical reactions.
Main Results:
- Seven patients experienced grade 2 hypersensitivity reactions within the first minute of dialysis.
- Significantly higher ETO concentrations were detected in the polyurethane potting of the dialyzers compared to the capillaries.
- Reactions were mitigated when extensive rinsing of the dialyzers was employed.
Conclusions:
- The acute hypersensitivity reactions are presumed to be related to residual ETO in the dialyzers.
- ETO concentration in the dialyzer potting may be a critical factor in reaction development.
- Further investigation is warranted, though other mechanisms cannot be entirely excluded.