Assessment of left ventricular function with 16- and 64-slice multi-detector computed tomography.
Suhny Abbara1, Benjamin J W Chow, Antonio J Pena
1Department of Radiology, Massachusetts General Hospital and Harvard Medical School, 165 Cambridge Street, Suite 400, Boston, MA 02114, United States. Sabbara@partners.org
The 64-slice MDCT accurately assesses left ventricular ejection fraction (LVEF) and dimensions, comparable to echocardiography and SPECT. This advanced imaging offers superior accuracy over 16-slice MDCT for cardiac function evaluation.
Area of Science:
- Cardiovascular Imaging
- Medical Diagnostics
- Radiology
Background:
- Accurate assessment of left ventricular ejection fraction (LVEF) is crucial for cardiac patient risk stratification and management.
- Imaging modalities providing both anatomical and functional data are highly desirable.
- Electrocardiographic (ECG) gated multi-detector computed tomographic (MDCT) imaging can assess LV function, but past systemic errors were linked to slow gantry rotation speeds.
Purpose of the Study:
- To evaluate the accuracy of ECG-gated MDCT in assessing LVEF and LV dimensions.
- To compare the performance of 16-slice and 64-slice MDCT scanners against echocardiography (ECHO) and Tc-99m gated single-photon-emission computed tomography (SPECT).
Main Methods:
- 306 patients underwent ECG-gated cardiac CT studies; 69 with available CT data and recent ECHO/SPECT were included.
- ECG-gated data sets were acquired using either 16-slice or 64-slice MDCT.
- Functional MDCT data were reconstructed in 10 cardiac phases and interpreted by blinded observers.
Main Results:
- The 64-slice MDCT showed very good correlation with combined ECHO/SPECT (r=0.86), ECHO (r=0.89), and SPECT (r=0.90) for LVEF.
- The 16-slice MDCT demonstrated fair correlation with combined ECHO/SPECT (r=0.56), poor with ECHO (r=0.26), and good with SPECT (r=0.76).
- LVEF values from 64-slice MDCT were similar to those from ECHO and SPECT.
Conclusions:
- 64-slice MDCT provides accurate LVEF and LV dimension assessments comparable to echocardiography and Tc-99m gated SPECT.
- The 64-slice MDCT (330 ms gantry rotation) demonstrates superior accuracy compared to 16-slice MDCT (420 ms gantry rotation) for cardiac function evaluation.
- ECG-gated 64-slice MDCT is a valuable tool for assessing cardiac function in patients.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Mitral Stenosis II: Clinical features and Diagnostic Tests
Imaging Studies for Cardiovascular System IV: CMRI