Related Experiment Video
Updated: Jul 11, 2026

Use of the Operant Orofacial Pain Assessment Device (OPAD) to Measure Changes in Nociceptive Behavior
Published on: June 10, 2013
A designed experiment for evaluation of the OPA method for cleaning studies of medical devices
Patrick J McCormick1, James J Kaiser, Michael J Schoene
1Bausch & Lomb, Inc., USA. pmccormick@bausch.com
Abstract:
This study reports the results of a designed experiment to evaluate the use of the o-phthalic dialdehyde (OPA) method for residual protein, with a testsoilderivedfrom blood, dehydrated hog mucin, and egg yolk, and its application in validating the automated cleaning of surgical instruments in a laboratory setting. The ruggedness and robustness of the OPA method was determined by means of a Plackett-Burman experimental design and assessed the automated cleaning of a microkeratome for use in patients undergoing lamellar resection of the cornea during refractive surgery. The outcome of the study indicates that the OPA method for residual protein is rugged and robust for the factors evaluated and is an effective means for validating the cleaning of complex medical devices under controlled laboratory conditions. The components of the microkeratome contaminated with the test soil were successfully cleaned in an automated washer-disinfector using a neutral pH detergent and the procedure as described.
Related Concept Videos
Experimental Designs
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Data Collection by Experiments
An example of the experimental method is a public clinical trial...
Methods of Medium Optimization