Cardiac output measurements with electrical velocimetry in patients undergoing CABG surgery: a comparison with
1Maribor Teaching Hospital, Department of Anaesthesiology, Intensive Care and Pain Management, Maribor, Slovenia.
Background And Objective:
The purpose was to study the agreement between cardiac output measurements with electrical velocimetry vs. intermittent thermodilution before and after coronary artery bypass graft surgery.
Methods:
Cardiac output was measured simultaneously with electrical velocimetry and intermittent thermodilution before and immediately after coronary artery bypass graft surgery, and in the intensive care unit. Measurements were performed in three different body positions. The results were analysed according to Bland and Altman.
Results:
The mean bias of all 150 paired measurements in 16 patients was 0.21 +/- 0.78 L min(-1), and the mean error was 40%. Before skin incision the mean bias was 0.04 +/- 0.41 L min(-1), and the mean error was 25%. After skin closure the mean bias was 0.57 +/- 0.92 L min(-1), and the mean error was 42%. In the intensive care unit the mean bias was 0.26 +/- 0.68 L min(-1), and the mean error was 32%.
Conclusions:
The agreement between cardiac output measurements with electrical velocimetry and intermittent thermodilution was clinically acceptable only before skin incision in coronary artery bypass graft surgery. The mean error was unacceptably high immediately after skin closure and was at a borderline level in the intensive care unit. Thus, the overall accuracy of cardiac output measurements with the electrical velocimetry technique during coronary artery bypass graft surgery is not clinically unacceptable.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be met...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...