Similarity between the suprasystolic wideband external pulse wave and the first derivative of the intra-arterial
R A Payne1, D Isnardi, P J D Andrews
1Clinical Pharmacology Unit, Centre for Cardiovascular Science, he University of Edinburgh, Edinburgh, UK.
Background:
Wideband external pulse (WEP) monitoring, using a broad bandwidth piezoelectric sensor located over the brachial artery under the distal edge of a sphygmomanometer cuff, can be used for evaluating the contour of the arterial pressure pulse wave. The pulse contour contains valuable information relating to cardiovascular function which may be of clinical use in addition to blood pressure measurements. The aim of this study was to compare the shape of the WEP signal during inflation of the cuff to suprasystolic pressure, with intra-arterial pressure waves, after the administration of vasoactive drugs.
Methods:
Radial intra-arterial and suprasystolic WEP waveforms were recorded in 11 healthy men (mean 23 yr) before and at the end of infusion of glyceryl trinitrate, angiotensin II, norepinephrine, and salbutamol. Waveform similarity was assessed by comparing the timing and pressure of incident and reflected waves and by root mean square error (RMSE).
Results:
The WEP signal was found to closely resemble the first derivative of intra-arterial pressure. The WEP signal could be used to derive an arterial pressure wave with minimal bias in the timing of incident [- 8 (18) ms, mean (SD)] and reflected [- 1 (24) ms] waves. Augmentation index was underestimated by WEP [- 7 (18)%]. WEP also provided a measure of compliance which correlated with pulse wave velocity (r = - 0.44). RMSE values after the administration of each of the four drugs mentioned earlier were 12.4 (3.8), 17.7 (5.0), 22.1 (11.7), and 28.9 (22.4) mm Hg, respectively. Changes in derived WEP signals were similar to those measured by arterial line with all drugs.
Conclusions:
The suprasystolic WEP signals can be used to derive arterial pressure waves which, although not identical, track changes in the intra-arterial pulse wave induced by vasoactive drugs.
Related Concept Videos
Pulse
The pulse serves as a clinical indicator...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring pulse
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation