Performance of heated wire humidifiers: an in vitro study
Paolo Pelosi1, Davide Chiumello, Paolo Severgnini
1Department of Environment, Health, and Safety, University of Insubria-Circolo and Fondazione Macchi Hospital, Varese, Italy.
Journal of Critical Care
|September 18, 2007
Summary
Heated wire humidifiers (HWHs) prevent circuit condensation. However, performance varies, with some HWHs delivering lower gas temperature and absolute humidity than expected.
Area of Science:
- Respiratory care
- Biomedical engineering
- Mechanical ventilation
Background:
- Heated humidifiers (HHs) are crucial for conditioning inspired gases during mechanical ventilation.
- Water condensation in breathing circuits can lead to complications like airway occlusion and infection.
- Heated wire humidifiers (HWHs) were developed to minimize or eliminate circuit condensation.
Purpose of the Study:
- To assess the performance of three HWHs (MR850, CONCHATHERM IV, DAR HC 2000) in terms of gas temperature and absolute humidity (AH).
- To compare the efficiency of HWHs against a standard heated humidifier (MR730 HH).
- To evaluate the impact of different temperature settings and gradients on HWH performance.
Main Methods:
- Evaluated three HWH models and one standard HH using a test lung.
- Measured gas temperature and AH at the Y-piece under varying ventilation volumes (5 and 15 L/min).
- Tested different temperature settings (35°C, 37°C, 39°C) and temperature gradients (-2°C, 0°C, +2°C).
Main Results:
- All HWHs successfully eliminated water condensation in the inspiratory circuit.
- MR850 and CONCHATHERM IV delivered lower gas temperature and AH compared to DAR HC 2000 and MR730 HH at a 0°C gradient.
- Increasing temperature gradients increased AH in all HWHs and gas temperature in the CONCHATHERM IV.
Conclusions:
- HWHs effectively prevent circuit condensation.
- Significant variations exist in the gas temperature and AH delivered by different HWH models.
- The actual gas temperature and AH delivered by HWHs may be lower than intended settings.