Objectives:
CorTemp is a wireless intestinal temperature monitoring system in the form of an ingestible pill and an external receiver. The aim of the study was to evaluate the system's accuracy and practicality during cardiac surgery.
Methods:
A repeat measures design using simultaneous temperature readings from the pulmonary artery (T (pa)), a nasopharyngeal thermometer (T (np)), skin thermometers (T (sk)) and the CorTemp system (T (in)), was conducted in 15 patients undergoing elective cardiac surgery under hypothermic conditions.
Results:
Only 67 % of patients' data was analysed and the statistical analysis of a total of 264 sets of readings showed a clinically significant temperature difference of T (in) compared to the other thermometers with limits of agreement between T (in) and T (pa), T (np) and T (sk) (+/- 0.35 to +/- 1.53 degrees C), (+/- 0.72 to +/- 1.63 degrees C) (+/- 0.40 to +/- 1.84 degrees C), respectively. The T (in) bias was significantly different from that of T (pa) ( P = 0.0023), T (np) ( P = 0.018) and T (sk) ( P = 0.0005) during rewarming. The T (in) rate of temperature change was also found to be significantly slower during the rewarming period.
Conclusions:
The significant temperature differences detected during rewarming urge caution regarding CorTemp use as an accurate estimator of brain temperature in cardiac surgery. Further studies are required to assess its potentially useful role as a body core and intestinal temperature monitoring system and as a useful adjunct in investigating bowel ischaemia aetiology in cardiac surgery.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...