Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Determination of Crystal Structures01:29

Determination of Crystal Structures

In the late 1800s, the revelation that light extended beyond visible wavelengths led to the discovery of X-rays by Wilhelm Roentgen. Recognized as high-energy electromagnetic radiation with short wavelengths, X-rays prompted exploration into their interaction with crystals. Max von Laue proposed in 1912 that the periodic arrangement of atoms, ions, or molecules in crystals would cause them to diffract X-rays, a hypothesis confirmed through experiments with copper sulfate and zinc sulfide...
Computed Tomography01:10

Computed Tomography

Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Comparison of a contrast-to-noise ratio-driven exposure control and a regular detector dose-driven exposure control in abdominal imaging in a clinical angiography system.

Medical physics·2021
Same author

Impact of a contrast-to-noise ratio driven and material specific exposure control on image quality and radiation exposure in angiography.

Physics in medicine and biology·2021
Same author

Performance of Photon-Counting Breast Computed Tomography, Digital Mammography, and Digital Breast Tomosynthesis in Evaluating Breast Specimens.

Academic radiology·2016
Same author

Metal artifact reduction for flat panel detector intravenous CT angiography in patients with intracranial metallic implants after endovascular and surgical treatment.

Journal of neurointerventional surgery·2015
Same author

Optical measurement of dimensional parameters of the breast with subjects in prone position.

Journal of medical imaging (Bellingham, Wash.)·2015
Same author

Impact of a new metal artefact reduction algorithm in the noninvasive follow-up of intracranial clips, coils, and stents with flat-panel angiographic CTA: initial results.

Neuroradiology·2013

Related Experiment Video

Updated: Jul 11, 2026

Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating
10:39

Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating

Published on: October 11, 2016

Efficiency of antiscatter grids for flat-detector CT.

Yiannis Kyriakou1, Willi Kalender

  • 1Institute of Medical Physics, University of Erlangen-Nürnberg, Henkestrasse 91, 91052 Erlangen, Germany. yiannis.kyriakou@imp.uni-erlangen.de

Physics in Medicine and Biology
|October 9, 2007
PubMed
Summary

Antiscatter grids in flat-panel detector CT (FD-CT) reduce artifacts but may not improve signal-to-noise ratio (SNR). For high scatter, grids can significantly reduce patient dose.

More Related Videos

Sample Preparation for In Situ Cryotomography of Mammalian Cells
08:17

Sample Preparation for In Situ Cryotomography of Mammalian Cells

Published on: December 15, 2023

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
06:28

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera

Published on: January 30, 2020

Related Experiment Videos

Last Updated: Jul 11, 2026

Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating
10:39

Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating

Published on: October 11, 2016

Sample Preparation for In Situ Cryotomography of Mammalian Cells
08:17

Sample Preparation for In Situ Cryotomography of Mammalian Cells

Published on: December 15, 2023

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
06:28

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera

Published on: January 30, 2020

Area of Science:

  • Medical Imaging
  • Radiological Physics

Background:

  • Flat-panel detector CT (FD-CT) offers large volume coverage but is prone to scatter artifacts.
  • Antiscatter grids are a common technique for scatter reduction in CT imaging.

Purpose of the Study:

  • To evaluate the scatter suppression efficiency and image quality improvement of three standard fluoroscopic antiscatter grids for FD-CT.
  • To assess the impact of antiscatter grids on signal-to-noise ratio (SNR) and artifacts.

Main Methods:

  • Simulations using a hybrid deterministic/Monte Carlo model and analytical grid transmission calculations.
  • Measurements of scatter-to-primary ratio (SPR) using an adapted collimator technique to validate simulations.
  • Evaluation of image quality metrics including SNR and artifact reduction.

Main Results:

  • Simulated and measured SPR values agreed within 10%.
  • Antiscatter grids significantly reduced cupping artifacts but did not generally improve SNR.
  • A trade-off exists between SNR and scatter reduction, quantified by the signal-to-noise improvement factor (SNR(if)).

Conclusions:

  • Antiscatter grids are effective in reducing scatter artifacts in FD-CT.
  • For high scatter conditions, grids can lead to dose reduction up to 50% while maintaining or improving image quality.
  • Optimization of exposure is necessary for low-to-medium scatter conditions to compensate for increased noise due to primary beam attenuation.