Related Experiment Video
Updated: Jul 10, 2026

12:25
C-arm-Free Simultaneous OLIF51 and Percutaneous Pedicle Screw Fixation in a Single Lateral Position
Published on: September 16, 2022
Interobserver reliability in evaluation of pedicle screw positions inserted with a modified technique
Ozgur Ozer1, Cagatay Ozturk, Burak Akesen
1Department of Orthopaedic Surgery, Faculty of Medicine, University of Uludağ, Bursa, Turkey.
Acta Orthopaedica Belgica
|October 18, 2007
Summary
This study evaluated a new pedicle screw insertion technique. While simple and reproducible, plain radiographs showed poor reliability for assessing screw position, unlike CT scans.
Area of Science:
- Spine surgery
- Orthopaedic surgery
- Surgical techniques
Background:
- Pedicle screw fixation is a common spinal stabilization technique.
- Accurate screw placement is crucial for surgical success and patient outcomes.
- Assessing pedicle screw position postoperatively typically involves imaging modalities.
Purpose of the Study:
- To evaluate a novel technique for pedicle screw insertion.
- To assess the reliability of plain radiographs and computed tomography (CT) scans in determining pedicle screw position.
- To analyze interobserver reliability in evaluating screw placement across different spinal regions and conditions.
Main Methods:
- A new pedicle screw insertion technique was applied to 201 screws in 27 patients.
- Screw positions were independently evaluated by two orthopaedic surgeons and two radiologists.
- Interobserver reliability was assessed for both plain radiographs and CT scans, stratified by spinal segment and condition.
Main Results:
- Malpositioned screw rates varied: 6.5%-32.8% on plain radiographs and 3.5%-6.5% on CT scans.
- Interobserver reliability was poor for plain radiographs and fair for CT scans.
- The new insertion technique was found to be simple and reproducible.
Conclusions:
- Plain radiographs alone are unreliable for assessing pedicle screw accuracy.
- CT scans offer better interobserver reliability for evaluating screw placement.
- Development of a standardized classification system is recommended to enhance assessment reliability.