Related Experiment Video
Updated: Jul 10, 2026

Fiber-optic Implantation for Chronic Optogenetic Stimulation of Brain Tissue
Published on: October 29, 2012
An implantable synthetic SPECT lesion: a bridge from phantom to reality
M Bret Abbott1, Lars R Furenlid, Don W Wilson
1Center for Gamma-Ray Imaging, Department of Radiology, University of Arizona, Tucson, AZ, USA.
Unlabelled:
Small-animal imaging systems are often characterized using phantoms, which may not predict performance in clinical applications. An implantable synthetic SPECT lesion would facilitate characterization of lesion detectability in a living animal.
Methods:
Anion-exchange columns with bed volumes of 100-300 nL were constructed from medical-grade polyvinyl chloride tubing and resin. The columns were tested in an excised mouse femur and implanted in the femur of a living mouse. Imaging was performed using a prototype dual-modality SPECT/CT system.
Results:
Activity of 7.4-22.2 MBq (0.2-0.6 mCi) localized within the synthetic lesion. The synthetic lesions were reused multiple times. Mice tolerated the implanted columns without complications for up to 8 wk.
Conclusion:
A reusable, synthetic SPECT lesion was constructed and implanted in the femur of a living mouse. The synthetic lesion is useful for the development of imaging schemes and for more realistically evaluating imaging-system performance in the context of a living animal.