Related Experiment Video
Updated: Jul 10, 2026

Semi-Automatic Graphical Tool for Measuring Coronary Artery Spatially Weighted Calcium Score from Gated Cardiac Computed Tomography Images
Published on: September 22, 2023
Improved signal-to-noise ratio in parallel coronary artery magnetic resonance angiography using graph cuts based
Gurmeet Singh1, Thanh Nguyen, Bryan Kressler
1School of Electrical & Computer Engineering, Cornell University, Ithaca, NY, USA. gurmeet@cornell.edu
Abstract:
High resolution 3D coronary artery MR angiography is time-consuming and can benefit from accelerated data acquisition provided by parallel imaging techniques without sacrificing spatial resolution. Currently, popular maximum likelihood based parallel imaging reconstruction techniques such as the SENSE algorithm offer this advantage at the cost of reduced signal-to-noise ratio (SNR). Maximum a posteriori (MAP) reconstruction techniques that incorporate globally smooth priors have been developed to recover this SNR loss, but they tend to blur sharp edges in the target image. The objective of this study is to demonstrate the feasibility of employing edge-preserving Markov random field priors in a MAP reconstruction framework, which can be solved efficiently using a graph cuts based optimization algorithm. The preliminary human study shows that our reconstruction provides significantly better SNR than the SENSE reconstruction performed by a commercially available scanner for navigator gated steady state free precession 3D coronary magnetic resonance angiography images (n = 4).