Multicenter study on finger systolic blood pressure test for diagnosis of vibration-induced white finger
Yoshiro Nasu1, Youichi Kurozawa, Yutaka Fujiwara
1Clinical Research Center for Hand-Arm Vibration Syndrome, Japanese Labor, Health and Welfare Organization, San-in Rosai Hospital, Yonago, Japan. nasu378@sanin-rosai-h.yonago.tottori.jp
Objective:
A multicenter study (six Rosai hospitals around Japan) was performed to investigate the diagnostic value of changes in finger systolic blood pressure (FSBP) after segmental local cooling for vibration-induced white finger (VWF).
Methods:
Subjects were 154 men without exposure to vibration and 135 men with occupational vibration exposure. They were classified into four groups: Group A, 154 unexposed control cases; Group B, 21 exposed cases without VWF; Group C, 31 cases with a history of VWF but without any signs of VWF within the last year; and Group D, 83 cases with active VWF within the last year. FSBP% measurements were taken at room temperatures of 23 +/- 1 and 21 +/- 1 degrees C, using a strain-gauge Digimatic 2000 plethysmograph (Medimatic).
Results:
At a room temperature of 23 +/- 1 degrees C, there was a significant difference between Groups A and D, and B and D. At a room temperature of 21 +/- 1 degrees C, there was a significant difference between Groups A and C, A and D, and B and D. The values in Group D were the lowest at both room temperatures. Assuming a cut-off value of 75% at 23 +/- 1 degrees C, the sensitivity and specificity were 65.2 and 87.5%, respectively. Assuming the same cut-off value at 21 +/- 1 degrees C, the sensitivity and specificity were 73.9 and 82.5%, respectively. These values were not too high. Most of the subjects with WVF in this study were retired and had not used vibratory tools for many years. The situation of the subjects may affect the results of the FSBP test. Our data did not confirm a difference in diagnostic accuracy between room temperatures of 23 +/- 1 and 21 +/- 1 degrees C.
Conclusions:
Our study showed that the sensitivity and specificity of the FSBP test with a cut-off value of 75% at 23 +/- 1 degrees C, were 65.2 and 87.5%, respectively, and at 21 +/- 1 degrees C, they were 73.9 and 82.5%, respectively.
Related Concept Videos
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure in the Leg
Preparation: