Related Experiment Video
Updated: Jul 10, 2026

11:51
An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
A simplified method to assess precision of fit between framework and supporting implants: a preliminary study
Angelo Calderini1, Carlo Maiorana, Giuliano Garlini
1Department of Oral Surgery, School of Dentistry, University of Milan, Italy. angelocalderini@tiscali.it
The International Journal of Oral & Maxillofacial Implants
|November 3, 2007
Summary
A new method using the OsseoCare device and torque-angle signature analysis can assess the precision of fit for screw-retained dental frameworks. This technique accurately evaluated frameworks made with CNC and CTiP methods, unlike traditional 1-piece casting.
Area of Science:
- Biomaterials Science
- Dental Implantology
- Prosthodontics
Background:
- Assessing the precision of fit between implant-supported prostheses and supporting implants is crucial for long-term success.
- Traditional methods for evaluating framework fit can be complex and time-consuming in clinical settings.
Purpose of the Study:
- To introduce a simplified clinical method for evaluating the precision of fit for screw-retained frameworks on dental implants.
- To assess the feasibility of using torque-angle signature analysis with the OsseoCare device for this purpose.
Main Methods:
- Three titanium frameworks were fabricated using distinct methods: 1-piece casting, computer numeric controlled (CNC) machining, and Cresco Ti Precision (CTiP).
- The OsseoCare device was employed to perform torque-angle signature analysis on the fabricated frameworks.
- Analysis focused on identifying tension within the framework-implant interface.
Main Results:
- Frameworks produced via CNC and CTiP methods exhibited torque-angle signature analysis results comparable to passively fitting reference models.
- The 1-piece cast framework did not yield results indicative of a passive fit.
- The OsseoCare device successfully generated tracings for analysis.
Conclusions:
- The OsseoCare device, in conjunction with torque-angle signature analysis, offers a feasible approach for the clinical assessment of framework fit.
- This pilot study suggests that CNC and CTiP fabrication methods result in superior framework fit compared to 1-piece casting.