Related Experiment Video
Updated: Jul 10, 2026

Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Effect of transvenous cardiac resynchronization therapy device implantation on cardiac troponin I release
Timucin Altin1, Omer Akyurek, Kutay Vurgun
1Ankara University School of Medicine, Department of Cardiology, Ankara, Turkey. alitimaltin@yahoo.com
Background:
Pacemaker and implantable cardioverter defibrillator (ICD) implantation increases cardiac troponin I (cTnI) levels which indicates myocardial injury. During implantation of a cardiac resynchronization therapy (CRT) device, balloon inflation for coronary sinus (CS) venogram, cannulation of CS side branch, and electrode advancement may interfere with CS drainage and, hence, may decrease the washout of toxic metabolites from the heart. Thus, CRT implantation may further increase cTnI levels. In this study, we investigated the effects of CRT implantation on cTnI release.
Methods:
We included 10 patients (mean age = 57 +/- 15 years) in whom a successful transvenous CRT system was implanted (CRT group). Twenty patients (mean age = 65 +/- 10 years) who underwent a transvenous pacemaker or ICD implantation were included as the control group. Blood samples for cTnI were drawn at baseline and at six, 12, 18, and 24 hours thereafter.
Results:
Baseline median cTnI levels were similar in CRT and control groups (0.03 ng/mL vs 0.02 ng/mL, respectively; P = 0.1). Postoperative cTnI levels during 24 hours were significantly higher in the CRT group (P < 0.05) by two-way repeated measures of analysis of variance. Post hoc analysis revealed that cTnI levels were higher at the 6th, 12th, 18th, and 24th hours compared to baseline levels (P < 0.001, P < 0.001, P < 0.01, and P < 0.01, respectively). There was a significant difference in the area under the curves (AUCs) of cTnI measurements (1.79 hr.ng/mL in the CRT group and 0.78 hr.ng/mL in the control group, P < 0.05).
Conclusion:
Postoperative cTnI levels were higher after CRT implantation than simple pacemaker/ICD implantation. This may be due to CS manipulation during CRT implantation.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019