Quantitative study of solvent and surface effects on analyte ionization in desorption ionization on silicon (DIOS)
1Department of Chemistry, North Carolina State University, Raleigh, North Carolina 27695-8204, USA.
Abstract:
Deuterated solvents and DIOS surfaces derivatized with different functional groups are used to investigate impacts of local chemical environment on analyte ionization. Both solvent molecules and surface functional groups are found to directly participate in analyte protonation in the condensed phase. The corresponding protonation effectiveness is quantitatively estimated based on the relative MS peak intensities of [M+2]+/[M+1]+. A direct correlation between ionization of triethylamine and the relative acidities of the surface and the solvent is evident. In addition, the proton donating effectiveness of a solvent is found to be related to its vapor pressure. Improved MS detection of small molecules via proper surface treatment and solvent selection is demonstrated.