Related Experiment Video
Updated: Jul 10, 2026

A Method for Systematic Electrochemical and Electrophysiological Evaluation of Neural Recording Electrodes
Published on: March 3, 2014
Suitability of electrosurgical analysers for routine maintenance testing of electrosurgery units - a technical note
M C Dale1, L H Tasker, J P McCarthy
1Clinical Engineering, Cardiff & Vale NHS Trust, Quadrant Centre, Cardiff Business Park, Llanishen, Cardiff, UK. Megan.Dale@CardiffandVale.wales.nhs.uk
Primary Objective:
To evaluate the use of electrosurgical analysers in testing power output and leakage current from an electrosurgery unit and compare this to the manufacturer recommendations for routine testing.
Method:
Two electrosurgical analysers were compared to reference measurements (carried out using non-inductive resistors, a current transformer and oscilloscope) over a range of tests described in IEC 60601-2-2: 1998 measuring power output and leakage currents in different conditions. The analysers used were Metron QA-ES and Fluke 454A.
Outcomes:
Both analysers gave similar results to the reference measurements for power output. The Metron QA-ES gave similar results to the reference measurements for leakage current testing; however the Fluke 454A gave substantially different results when used as described in the manual.
Conclusions:
Electrosurgical analysers can be a valuable tool in the workshop, enabling rapid, accurate testing of electrosurgery equipment without needing additional equipment and setting up times. Not all analysers can perform all the tests that may be needed and in some cases the accuracy of the results is questionable. Users must be certain of the capabilities and limitations of the analyser before making decisions based on the results.
More Related Videos
04:48Development of a Low-cost Epimysial Electromyography Electrode: A Simplified Workflow for Fabrication and Testing
Published on: April 12, 2024
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
Related Concept Videos
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in the...
Electrogravimetric Analysis: Overview
To test the completeness of the...