Objective:
Development of a composite material phantom, comprised of polyvinyl alcohol cryogel (PVA-C) and an agarose additive, to effectively mimic the magnetic resonance relaxation times (T1 and T2) of neonatal white matter (WM) and gray matter (GM) at 3.0 T.
Materials And Methods:
Samples of PVA-C with and without agarose were prepared with 1 cycle of freezing/thawing. Measurements of T1 and T2, at 3.0 T, were performed on the samples at temperatures ranging from 20 degrees C to 40 degrees C.
Results:
A sample temperature of 40 degrees C was required to achieve a T1 value sufficiently long to represent neonatal WM. At this temperature, neonatal WM relaxation times required 3% PVA-C with 0.3% agarose, whereas gray matter relaxation times required 8% PVA-C with 1.4% agarose.
Conclusions:
By adjusting the sample temperature, polyvinyl alcohol concentration, and agarose concentration, the relaxation times of neonatal brain tissues can be obtained using this composite material.