A comparative study of the effect of hydrogen peroxide versus normal saline on the strength of the bone-cement
S R Guerin1, R MacNiochaill, P O'Reilly
1Trinity Centre for Bioengineering, Department of Mechanical and Manufacturing Engineering, Trinity College, Dublin, Ireland.
Abstract:
Hydrogen peroxide has been used for decades as an effervescent haemostatic agent in arthroplasty. Recently it has been shown to adversely affect the material properties of PMMA. We aim to assess whether any such deleterious effects are demonstrated in an experimental model which mimics the clinical use of hydrogen peroxide. Matched pairs of cancellous bone samples were treated with a swab soaked in either saline or a 3% solution of hydrogen peroxide, prior to manufacture of cement-bone constructs using Palacos or Simplex cement. Thirty pairs were then compared by subjecting them to a torsional shear force until failure and a further thirty pairs were tested to failure in tension. There was no significant difference between the mean torques to failure for the Palacos-peroxide group versus the Palacos-saline group, or the Simplex-peroxide versus the Simplex-saline group (p=0.31 and 0.71 respectively). Similarly there was no significant difference between the mean tension loads to failure for the Palacos-peroxide group versus the Palacos-saline group, and the Simplex-peroxide versus the Simplex-saline group (p=0.79 and 0.23 respectively). We conclude that the use of hydrogen peroxide as an effervescent haemostatic agent has no detrimental effect on the mechanical integrity of the bone-cement interface when compared to normal saline.
Insights
Hydrogen peroxide, used as a haemostatic agent in arthroplasty, does not negatively impact the mechanical strength of bone-cement interfaces. This study found no significant differences when comparing hydrogen peroxide to saline treatments.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Surgical Hemostasis
Background:
- Hydrogen peroxide is a common effervescent hemostatic agent in arthroplasty.
- Recent studies suggest potential adverse effects of hydrogen peroxide on polymethyl methacrylate (PMMA) properties.
Purpose of the Study:
- To evaluate the mechanical integrity of the bone-cement interface after using hydrogen peroxide as a hemostatic agent in an experimental model.
- To determine if hydrogen peroxide adversely affects cement-bone bond strength compared to saline.
Main Methods:
- Cancellous bone samples were treated with either 3% hydrogen peroxide or saline-soaked swabs.
- Bone-cement constructs were created using Palacos or Simplex cement.
- Samples were subjected to torsional shear and tensile testing to failure.
Main Results:
- No significant differences in mean torque to failure were observed between hydrogen peroxide and saline groups for both Palacos (p=0.31) and Simplex (p=0.71) cements.
- No significant differences in mean tension load to failure were found between hydrogen peroxide and saline groups for both Palacos (p=0.79) and Simplex (p=0.23) cements.
Conclusions:
- The use of hydrogen peroxide as an effervescent hemostatic agent does not compromise the mechanical integrity of the bone-cement interface.
- Clinical application of hydrogen peroxide in arthroplasty appears safe regarding bone-cement bond strength.
Related Concept Videos
Strength and Heat of Hydration
The heat of hydration for each cement compound is significant; for instance, tricalcium aluminate (C3A) and...
Testing Water Quality
Effect of Sea Water on Concrete
Concrete in areas between tide marks, which undergo...
Porosity in Cement Paste
The balance of water to cement in the mix is critical—it...
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in ASTM C...
Pore Size Distribution
Adequate...