Misconceptions in reporting oxygen saturation
John Toffaletti1, Willem G Zijlstra
1From the *Department of Pathology, Duke University Medical Center, North Carolina; and †Department of Pediatrics, University Medical Center Groningen, Groningen, The Netherlands.
Background:
We describe some misconceptions that have become common practice in reporting blood gas and cooximetry results. In 1980, oxygen saturation was incorrectly redefined in a report of a new instrument for analysis of hemoglobin (Hb) derivatives. Oxygen saturation (sO2) was redefined as the ratio of oxyhemoglobin (O2Hb) to total Hb instead of the ratio of O2Hb to active Hb (O2Hb + desoxyhemoglobin). In addition, the new terms "functional saturation" and "fractional saturation" were introduced. Since the new parameter was implemented in a widely used cooximeter, its use is now widespread and has caused misunderstandings.
Methods:
In this report, we review the development of the definitions and measurements of sO2 and related quantities and contend that the misconceptions should be resolved by standardizing instrument read-outs and clinical reports, so that sO2, defined as the ratio of O2Hb to active Hb, should replace FO2Hb and be reported along with the total Hb concentration and the common dyshemoglobin fractions (%CO-Hb and % methemoglobin [metHb]).
Results:
The redefinition of sO2 as the %O2Hb or FO2Hb did not address the confusion that might result from interchanging these two often-similar but different terms. The term fractional saturation is an inappropriate terminology and lacks clear physiological meaning. We see frequent cases of confusion: (a) the difference between the sO2 in pulse oximetry and the FO2Hb in cooximetry is called the "pulse oximeter gap;" (b) sO2 results are described as "method dependent;" and (c) reference ranges for these terms are substituted.
Conclusions:
Although either parameter could be used by clinicians who fully understand the relatively simple difference between these parameters, we find clear evidence that there is widespread confusion of these terms, even among experts in the field. Standardization of the reporting format would help, and instrument manufacturers could contribute by standardizing the reporting format for cooximetry results.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
09:04Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...