Lessons learned with noninvasive cardiac monitoring for acute rejection
M Schweiger1, A Wasler, G Prenner
1Medical University Graz, University f. Surgery, Graz, Austria.
Introduction:
Computerized Heart Allograft Rejection Monitoring (CHARM), used for noninvasive rejection monitoring in heart transplant recipients, is based on the analysis of ventricular evoked response (VER) signals. This study evaluated the prognostic validity of the TslewC, a parameter extrapolated from the VER.
Methods:
During orthotopic heart transplantation (OHT) 2 unipolar, fractally coated, screw-in leads implanted epimyocardially were connected to a telemetric pacemaker. Recordings of IEGMs were performed routinely at hospital and at outpatient visits. Data processing yielded trend curves. TslewC was calculated from the tangent of VER. One hundred five patients divided into survivors and nonsurvivors, were compared using a two-tailed Student's t test.
Results:
In the final follow-up a significant lower TslewC was observed among patients in the nonsurvivor compared with the other group (P<.001). Tests to find an optimal prognostic threshold of the TslewC yielded the value of 26 mV.
Conclusion:
TslewC functioned as a prognostic factor after OHT. Further studies must provide a prognostic threshold to avoid patient visits all 4 weeks. Patients would only have to be admitted to the hospital if the TslewC was under this prognostic threshold.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiac Catheterization IV: Nursing Management
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Kidney Transplant III: Nursing Management
Cardiac Catheterization I: Pre-Procedure Overview