Automatic blood pressure measurement: the oscillometric waveform shape is a potential contributor to differences
John N Amoore1, Yann Lemesre, Ian C Murray
1Department of Medical Physics, Royal Infirmary of Edinburgh, Edinburgh, UK. john.amoore@luht.scot.nhs.uk
Objective:
To explore the differences between oscillometric and auscultatory measurements.
Method:
From a simulator evaluation of a non-invasive blood pressure (NIBP) device regenerating 242 oscillometric blood pressure waveforms from 124 subjects, 10 waveforms were selected based on the differences between the NIBP (oscillometric) and auscultatory pressure measurements. Two waveforms were selected for each of five criteria: systolic over and underestimation; diastolic over and underestimation; and close agreement for both systolic and diastolic pressures. The 10 waveforms were presented to seven different devices and the oscillometric-auscultatory pressure differences were compared between devices and with the oscillometric waveform shapes.
Results:
Consistent patterns of waveform-dependent over and underestimation of systolic and diastolic pressures were shown for all seven devices. The mean and standard deviation, for all devices, of oscillometric-auscultatory pressure differences were: for the systolic overestimated waveforms, 36 +/- 28/-6 +/- 3 and 23 +/- 2/-1 +/- 3 mmHg (systolic/diastolic differences); for systolic underestimated waveforms, -21 +/- 5/-4 +/- 3 and -11 +/- 4/-3 +/- 3 mmHg; for diastolic overestimated waveforms, 3 +/- 4/12 +/- 5 and 17 +/- 6/10 +/- 2 mmHg; for diastolic underestimated waveforms, 1 +/- 4/-22 +/- 4 and -9 +/- 6/-29 +/- 4 mmHg; and for the two waveforms with good agreement, 0 +/- 6/0 +/- 3 and -2 +/- 4/-4 +/- 3 mmHg. Waveforms for which devices showed good oscillometric and auscultatory agreement had smooth envelopes with clearly defined peaks, compared with the broader plateau and complex shapes of those waveforms for which devices over or underestimated pressures.
Conclusion:
By increasing the understanding of the characteristics and limitations of the oscillometric method and the effects of waveform shape on pressure measurements, simulator evaluation should lead to improvements in NIBP devices.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: