Related Experiment Video
Updated: Jul 9, 2026

Snap Chip for Cross-reactivity-free and Spotter-free Multiplexed Sandwich Immunoassays
Published on: November 13, 2017
Competitive immunoassays for simultaneous detection of metabolites and proteins using micromosaic patterning
Brian M Murphy1, Xinya He, David Dandy
1Department of Chemistry, and Department of Chemical and Biochemical Engineering, Colorado State University, Fort Collins, Colorado 80523, USA.
Abstract:
New high-throughput immunoassay methods for rapid point-of-care diagnostic applications represent an unmet need and current focus of numerous innovative methods. We report a new micromosaic competitive immunoassay developed for the analysis of the thyroid hormone thyroxine (T4), inflammation biomarker C-reactive protein (CRP), and the oxidative damage marker 3-nitrotyrosine (BSA-3NT) on a silicon nitride substrate. To demonstrate the versatility of the method, both direct and indirect format competitive immunoassays were developed and could be applied simultaneously for single samples. Signals from standard solutions were fit to a logistic equation, allowing simultaneous detection of T4 (7.7-257.2 nM), CRP (0.3-4.2 microg/mL), and BSA-3NT (0.03-22.3 microg/mL). Total assay time including sample introduction, washing, and fluorescence measurement was less than 45 min. Dissociation constants for affinity pairs in the system have been estimated using regression. This proof-of-concept experiment shows that both small and macromolecular biomarkers can be quantified from a single sample using the method and suggests that groups of clinically related analytes may be analyzed by competitive micromosaic immunoassay techniques.