Method validation for the determination of lead in raw cow's milk by electrothermal atomic absorption spectrometry
Deniz Yurtsever Sarica1, Ali Rehber Türker
1TUBITAK/ATAL Scientific and Technological Research Council of Turkey/Ankara Test and Analysis Laboratory, Beşevler, Ankara 06530, Turkey.
Abstract:
In this study, lead in raw cow's milk has been determined by validated electrothermal atomic absorption spectrometry (ETAAS) with Zeeman-effect background correction. Maximum pyrolysis and optimum atomization temperatures of lead were determined in the presence of modifiers. Pd + Mg(NO3)2 has been found a powerful modifier mixture for the determination of lead. The analytical parameters of the method such as limit of detection, limit of quantification and the effect of interfering ions have been investigated. The detection limit (3sigma) achieved by the method was calculated to be 0.62 ng/mL for Pb. Repeatability of the method evaluated as the relative standard deviation of 16-17 replicates using 5 ng/mL, under optimum experimental conditions were about 1.5% for synthetic sample solution and about 3% for real sample (N = 5). The described method has been validated by analyzing certified reference material (BCR-CRM 150) and by comparing the results with those obtained by inductively coupled plasma mass spectrometry (ICP-MS). The validated method was applied to raw cow's milk samples produced in 7 different regions of Turkey in 2003-2004. Raw cow's milk contained a mean (range) of 31.4 (2.5 - 313) microg kg(-1) lead with a relative error below 2%.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
10:48Electrochemical Preparation of Poly(3,4-Ethylenedioxythiophene) Layers on Gold Microelectrodes for Uric Acid-Sensing Applications
Published on: July 28, 2021
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Data Validation
Key parameters for method validation include:
Atomic Emission Spectroscopy: Lab
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
Atomic Emission Spectroscopy: Overview
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...