Precision of In Vivo Quantitative Tooth Wear Measurement Using Intra-Oral Scans
Published on: July 12, 2022
Precision of wear measurement using the shadowgraph technique
J L Cunningham1, I Bisbinas, R Greenwood
1Department of Orthopaedic Surgery, University of Bristol, and Research and Development Support Unit, United Bristol Healthcare Trust, UK. j.l.cunningham@bath.ac.uk
The shadowgraph technique offers an accessible method for measuring wear in acetabular cups. While linear wear measurement is precise, observer variability significantly impacts results more than mould variations.
Area of Science:
- Orthopedics
- Biomaterials Science
- Medical Device Analysis
Background:
- Acetabular cup wear is a critical factor in total hip replacement longevity.
- Accurate wear measurement is essential for evaluating implant performance and failure modes.
- The shadowgraph technique presents a potentially simple method for assessing wear in explanted devices.
Purpose of the Study:
- To evaluate the measurement precision of the shadowgraph technique for acetabular cups.
- To determine the influence of different observers on shadowgraph measurements.
- To assess the impact of using multiple moulds on the accuracy of wear quantification.
Main Methods:
- Utilized the shadowgraph technique on moulds derived from explanted acetabular cups.
- Measured linear wear and wear angle to calculate volumetric wear.
- Assessed measurement precision using coefficient of variation (CV).
- Analyzed the influence of observer variability and mould differences on results.
Main Results:
- Linear wear measurement demonstrated high precision (CV = 1.49%).
- Wear angle measurement showed lower precision (CV = 8.18%).
- Both observer and mould significantly affected measurement outcomes.
- Observer influence was substantially greater than mould influence.
Conclusions:
- The shadowgraph technique is a viable, low-cost method for acetabular cup wear assessment.
- Linear wear is more reliably measured than wear angle using this technique.
- Standardization of observer protocols is crucial to minimize variability and enhance the reliability of shadowgraph wear measurements in orthopedic research.