The USP Performance Verification Test, Part II: collaborative study of USP's Lot P Prednisone Tablets
Maria Glasgow1, Shawn Dressman, William Brown
1United States Pharmacopeia, Rockville, MD 20852, USA.
Purpose:
Periodic performance verification testing (PVT) is used by laboratories to assess and demonstrate proficiency and for other purposes as well. For dissolution, the PVT is specified in the US Pharmacopeia General Chapter Dissolution <711> under the title Apparatus Suitability Test. For Apparatus 1 and 2, USP provides two reference standard tablets for this purpose. For each new lot of these reference standards, USP conducts a collaborative study.
Methods:
For new USP Lot P Prednisone Tablets, 28 collaborating laboratories provided data. The study was conducted with three sets of tablets: Lot O open label, Lot O blinded, and Lot P blinded. The blinded Lot O data were used for apparatus suitability testing.
Results:
Acceptance limits were determined after dropping data due to failure of apparatus suitability, identification of data as unusual on control charts, or protocol violations.
Conclusions:
Results yielded acceptance criteria of (47, 82) for Apparatus 1 and (37, 70) for Apparatus 2. Results generally were similar for Lot P compared to results from Lot O except that the average percent dissolved for Lot P is greater than for Lot O with Apparatus 2.
Related Concept Videos
Drug Product Performance: In Vitro–In Vivo Correlation
In Vitro Drug Dissolution: Compendial Testing Models II
Pharmaceutical Alternatives: Polymorphic Form-Related and Particle Size-Related Therapeutic Nonequivalence
Pharmaceutical Alternatives: Stability-Related Therapeutic Nonequivalence
Clinically Relevant Drug Product Specifications: Methods of Establishment
Drug Dissolution: Requirements and Profile Comparison