Related Experiment Video
Updated: Jul 8, 2026

Assessing Endothelial Vasodilator Function with the Endo-PAT 2000
Published on: October 15, 2010
A preliminary evaluation of the mean arterial pressure as measured by cuff oscillometry
Harold Smulyan1, Paul R Sheehe, Michel E Safar
1Department of Medicine, Upstate Medical University, State University of New York, Syracuse, New York, USA. smulyanh@upstate.edu
Background:
The brachial artery (BA) mean blood pressure (MBP) is now readily available using the oscillometric technique. In contrast to the auscultatory method where MBP is calculated from the systolic (SBP) and diastolic blood pressure (DBP), oscillometric MBP is measured separately from either SBP or DBP. Because the peripheral MBP is free of amplification, it is nearly the same throughout the entire arterial tree and could represent the corresponding aortic pressure. The oscillometric brachial MBP could therefore serve as a surrogate for aortic MBP and might be a valuable non-invasive risk predictor.
Methods:
This study compares the oscillometric BA pressures with simultaneously and directly recorded aortic pressures in 100 patients.
Results:
These results show that, over a wide range of cuff pressures, the oscillometric MBP, whether alone or with age in multiple regression, predicts aortic pressure better than the SBP or DBP do, with a better correlation coefficient (r = +0.91), low aortic-cuff MBP difference (-0.79 mm Hg) and the lowest s.d. of the individual differences (+7.2 mm Hg).
Conclusions:
These results are preliminary and need to be confirmed by larger studies. If confirmed, the predicted aortic pressures should be calculated and displayed by the oscillometric BP devices, the goal being to develop better non-invasive cardiovascular (CV) risk predictors.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.