Related Experiment Video
Updated: Jul 8, 2026

Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
Axillary and rectal temperature measurements poorly agree in newborn infants
P C E Hissink Muller1, L H van Berkel, A J de Beaufort
1Neonatal Unit, Juliana Children's Hospital, The Hague, The Netherlands. p.hissinkmuller@lumc.nl
Aim:
Evaluation of the agreement between axillary temperature measurements and rectal temperature measurements in neonates.
Methods:
Rectal and axillary body temperatures were simultaneously measured for 3 min in 33 neonates (gestational age 25-42 weeks, weight 840-4,005 g). Two investigators performed paired measurements, one in each neonate. A single type of thermometer was used in this study: one thermometer for each rectal and another thermometer for each axillary measurement. The Bland-Altman method was used (95% 'limits of agreement': mean +/- 2 SD) to determine the level of agreement between axillary and rectal measurements.
Results:
The axillary temperature was significantly lower than the rectal temperature (mean +/- SD 0.27 +/- 0.20 degrees C, p < 0.05). The '95% limits of agreement' ranged from -0.13 to +0.67 degrees C. Increasing postnatal age (days) showed a significant increase in temperature difference (rectal minus axillary; r = 0.54; p < 0.05).
Conclusions:
The mean difference between axillary and rectal temperature shows a wide variation. Axillary temperature measurements cannot be used interchangeably with rectal measurements in neonates.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4: