Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
[Evaluation of cardiac function with 32DAS MDCT--fundamental examination]
Sumito Maruyama1, Toshiyuki Takahashi, Yoshimasa Takahashi
1Central Radiology Department, Showa University Fujigaoka Hospital.
Image acquisition settings impact cardiac ejection fraction (EF) calculations. Coronary CT (computed tomography) derived EF values showed variations compared to left ventriculography (LVG) in phantom and patient studies.
Area of Science:
- Cardiovascular Imaging
- Medical Physics
Background:
- Accurate assessment of cardiac function, particularly ejection fraction (EF), is crucial for diagnosing and managing cardiovascular diseases.
- Computed tomography (CT) offers a non-invasive method for cardiac imaging, but its accuracy in quantifying EF requires validation against established techniques.
Purpose of the Study:
- To investigate the influence of varying imaging parameters on ejection fraction (EF) measurements using 32-detector row multi-slice CT (MDCT).
- To compare EF values derived from coronary CT angiography (CCTA) with those obtained from left ventriculography (LVG) in a clinical setting.
Main Methods:
- Cardiac phantom studies were conducted with manipulated pulse counts, X-ray tube revolving speed, image reconstruction modes, and beam pitch.
- Clinical data from four patients who underwent both CCTA and LVG were analyzed for comparative EF assessment.
Main Results:
- Phantom image analysis revealed that coronary CT-derived EF values were generally smaller than LVG-derived values.
- End-diastolic volume (EDV) and end-systolic volume (ESV) measured by coronary CT were smaller than those measured by LVG.
- Conversely, EF values measured by coronary CT were found to be larger than those measured by LVG in some evaluations, with a maximal difference of approximately 10% relative to LVG values.
Conclusions:
- Image acquisition conditions significantly affect the calculated ejection fraction (EF) values in cardiac CT.
- While coronary CT provides valuable functional information, discrepancies exist compared to LVG, necessitating careful consideration of imaging protocols and potential biases in EF quantification.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System IV: CMRI
Mitral Stenosis II: Clinical features and Diagnostic Tests
Dysrhythmias V: Evaluating Dysrhythmias
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...