Calibration drift in a laboratory high purity germanium detector spectrometry system
1The Radiological Health Engineering Laboratory, Department of Nuclear Engineering and Radiological Sciences, The University of Michigan, 2355 Bonisteel Boulevard, 1906 Cooley Building, Ann Arbor, MI 48109-2104, USA.
Abstract:
For unknown radionuclide identification, it is important that a high purity germanium (HPGe) spectrometry system be calibrated correctly for energy. The energy calibration of an HPGe spectrometry system will drift over time due to a variety of factors including the ambient temperature, the line voltage applied to the system, variation in the electronics, and other possible influences. In order to better understand the nature of this energy calibration drift, calibration spectra were collected over a period of several months from a laboratory HPGe spectrometry system. System parameters, including detector voltage, amplifier gain, and preamplifier gain, were not deliberately modified during the course of the experiment. The system was calibrated routinely over the 90 days, and the results of the calibrations were compared in order to assess the drift in the energy calibration of the detector over time. The analysis of a 36% high purity germanium system demonstrated the energy calibration drifted an average of 0.014 keV d(-1) to 0.041 keV d(-1) depending upon energy. At 1,332 keV, one day after calibration, it was shown that up to half of the total error in energy calibration was as a result of calibration drift.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Uncertainty in Measurement: Reading Instruments
Gas Chromatography: Types of Detectors-II