Related Experiment Videos
A sensitive fluorimetric assay for serum angiotensin-converting enzyme.
American Journal of Clinical Pathology
|August 1, 1976
Summary
A new assay accurately measures angiotensin-converting enzyme (ACE) in serum. This rapid and sensitive method aids in diagnosing conditions like sarcoidosis.
Area of Science:
- Biochemistry
- Enzyme Assays
- Clinical Diagnostics
Background:
- Angiotensin-converting enzyme (ACE) plays a crucial role in the renin-angiotensin system.
- Accurate measurement of ACE activity in serum is important for diagnosing and managing certain diseases.
Purpose of the Study:
- To develop and validate a simple, rapid, sensitive, and reproducible assay for quantifying angiotensin-converting enzyme (ACE) activity in untreated human serum.
Main Methods:
- The assay utilizes the substrate hippuryl-L-histidyl-L-leucine, measuring the product hippurate spectrofluorimetrically after reaction with ophthaldialdehyde.
- Optimized conditions include 5 mM substrate in 0.1 M potassium phosphate buffer (pH 8.3) with 0.3 M NaCl.
- Enzyme kinetics were determined, with a Km of 1.33 mM for the substrate.
Main Results:
- The assay demonstrated chloride dependence and characteristic inhibition/activation patterns consistent with ACE.
- The mean serum ACE level in 58 healthy individuals (age range 19-57) was determined to be 32.2 ± 1.30 nmol/min/ml serum (SD = 9.87).
Conclusions:
- The developed assay provides a reliable method for measuring serum ACE activity.
- This assay is valuable for the diagnosis and management of sarcoidosis and may find broader clinical applications.