Prediction of perceptible artifacts in JPEG 2000-compressed chest CT images using mathematical and perceptual quality
Bohyoung Kim1, Kyoung Ho Lee, Kil Joong Kim
1Department of Radiology, Seoul National University Bundang Hospital, 300 Gumi-dong, Bundang-gu, Seongnam-si, Gyeonggi-do, Seoul 463-707, Korea.
Objective:
The objective of our study was to determine whether peak signal-to-noise ratio (PSNR) and a perceptual quality metric (High-Dynamic Range Visual Difference Predictor [HDR-VDP]) can predict the presence of perceptible artifacts in Joint Photographic Experts Group (JPEG) 2000-compressed chest CT images.
Materials And Methods:
One hundred chest CT images were compressed to 5:1, 8:1, 10:1, and 15:1. Five radiologists determined if the original and compressed images were identical (negative response) or different (positive response). The correlation between the results for each metric and the number of readers with positive responses was evaluated using Spearman's rank correlation test. Using the pooled readers' responses as the reference standard, we performed receiver operating characteristic (ROC) analysis to determine the cutoff values balancing sensitivity and specificity and yielding 100% sensitivity in each metric. These cutoff values were then used to estimate the visually lossless thresholds for the compressions for the 100 original images, and the accuracy of the estimates of two metrics was compared (McNemar test).
Results:
The correlation coefficients were -0.918 and 0.925 for PSNR and the HDR-VDP, respectively. The areas under the ROC curves for the two metrics were 0.983 and 0.984, respectively (p = 0.11). The PSNR and HDR-VDP accurately predicted the visually lossless threshold for 69% and 72% of the 100 images (p = 0.68), respectively, at the cutoff values balancing sensitivity and specificity and for 43% and 47% (p = 0.22), respectively, at the cutoff values reaching 100% sensitivity.
Conclusion:
Both metrics are promising in predicting the perceptible compression artifacts and therefore can potentially be used to estimate the visually lossless threshold.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
08:05Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
Published on: December 19, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...