The IFCC Reference Measurement System for HbA1c: a 6-year progress report.
Cas Weykamp1, W Garry John, Andrea Mosca
1Queen Beatrix Hospital, Winterswijk, The Netherlands. c.w.weykamp@skbwinterswijk.nl
Clinical Chemistry
|January 29, 2008
Summary
The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Reference Measurement System for hemoglobin A1c (HbA1c) demonstrates stability and continuity. Periodic evaluations confirm the reliability of this crucial system for accurate diabetes monitoring.
Area of Science:
- Clinical Chemistry
- Metrology
- Diabetes Mellitus Diagnostics
Background:
- The IFCC Reference Measurement System for hemoglobin A1c (HbA1c) is established with metrologic traceability.
- A network of 14 international reference laboratories supports the IFCC-RM.
- This study reports on 12 intercomparison studies assessing the IFCC-RM's performance.
Purpose of the Study:
- To evaluate the performance and stability of the IFCC Reference Measurement System for HbA1c.
- To assess the long-term continuity and reliability of HbA1c measurements.
- To compare the IFCC-RM with designated comparison methods (DCMs).
Main Methods:
- Utilized unknown and known samples, calibrators, and calibration sets for intercomparison studies.
- Measured unknown samples using the IFCC-RM and compared with DCMs (NGSP, JDS/JSCC, Mono-S).
- Investigated the stability of the Master Equation (ME) relating IFCC-RM and DCMs.
Main Results:
- 95 out of 105 IFCC-RM data sets were approved, indicating overall system validity.
- Trend analysis of MEs showed minimal, non-significant changes for NGSP and Mono-S, but a significant change for JDS/JSCC.
- Long-term performance showed no systematic change; mass spectrometry and capillary electrophoresis yielded comparable mean HbA1c, with CE showing better reproducibility. One calibrator batch was unapproved, but stored calibrators were stable.
Conclusions:
- The IFCC Reference Measurement System for HbA1c is sound, ensuring continuity and stability.
- The system provides a reliable analytical anchor for HbA1c measurements.
- Ongoing intercomparison studies are essential for maintaining the integrity of the IFCC-RM.