Atomic Absorption Spectroscopy: Interference01:25

Atomic Absorption Spectroscopy: Interference

Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Emission Spectroscopy: Interference01:30

Atomic Emission Spectroscopy: Interference

In atomic emission spectroscopy (AES), high-temperature atomizers excite a broad range of elements and molecules that generate complex emissions from sources such as oxides, hydroxides, and flame combustion products in the flame or plasma. Several strategies can be employed to minimize spectral interferences caused by overlapping emission lines or bands. These include increasing instrument resolution, choosing alternative emission lines, optimally placing the detector in low-background regions,...
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations01:08

IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations

Identical bonds within a polyatomic group can stretch symmetrically (in-phase) or asymmetrically (out-of-phase). Similar to hydrogen bonding, these vibrations also influence the shape of the IR peak. Generally, asymmetric stretching frequencies are higher than symmetric stretching frequencies. For example, primary amines exhibit two distinct IR peaks between 3300–3500 cm−1 corresponding to the symmetric and asymmetric N-H stretching, while secondary amines exhibit a single stretching vibration...
Atomic Absorption Spectroscopy: Radiation and Light Sources01:13

Atomic Absorption Spectroscopy: Radiation and Light Sources

Atomic absorption spectroscopy (AAS) relies on the Beer-Lambert law, which requires that the radiation source emits a narrow range of wavelengths to match the absorption characteristics of the analyte atom. The primary criteria for choosing an appropriate radiation source in AAS is to provide a precise and intense emission at specific wavelengths that will allow accurate detection of the analyte.
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Molecular Spectroscopy: Absorption and Emission01:14

Molecular Spectroscopy: Absorption and Emission

Molecules possess discrete energy levels called quantum states. Unlike atoms, which have simpler energy levels, molecules possess additional rotational and vibrational energy levels. Each energy level is separated by an energy gap, with the gaps between adjacent electronic, vibrational, and rotational levels varying significantly. The three types of energy levels in a diatomic molecule are shown in Figure 1.
Atomic Absorption Spectroscopy: Overview01:27

Atomic Absorption Spectroscopy: Overview

Atomic absorption spectroscopy (AAS) is a technique used to analyze elements by measuring electromagnetic radiation (EMR) absorbed by atoms, which causes them to transition to a higher-energy orbit. The most crucial step in AAS is atomization, where the analyte is converted into gas-phase atoms, typically through a flame or furnace. Some of these atoms become thermally excited in the flame, while most remain in the ground state.
When irradiated by EMR of a particular wavelength, these...