Related Experiment Video
Updated: Jul 7, 2026

Infrared Degenerate Four-wave Mixing with Upconversion Detection for Quantitative Gas Sensing
Published on: March 22, 2019
Trace gas detection by laser intracavity photothermal spectroscopy
1Brookhaven National Laboratory, Department of Applied Science, Environmental Chemistry Division, Upton, New York 11973, USA.
Abstract:
A novel laser intracavity photothermal detector is described. In this scheme, sample absorption of the pump laser power takes place within the cavity of a probe He-Ne laser causing modulation in the gain and in turn the output power. Comparison of this intracavity detector with two other photothermal techniques, namely, phase fluctuation optical heterodyne spectroscopy and thermal beam deflection, is made in terms of practicality and sensitivity. For in situ measurements, sensitivity of 0.5 x 10(-7) cm(-1) for a probe length of 3 cm has been achieved.
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II