Related Experiment Video
Updated: Jul 7, 2026

Mass Spectrometric Analysis of Glycosphingolipid Antigens
Published on: April 16, 2013
Simultaneous quantification of GM1 and GM2 gangliosides by isotope dilution tandem mass spectrometry
Jianghong Gu1, Cynthia J Tifft, Steven J Soldin
1Department of Laboratory Medicine, Children's National Medical Center, 111 Michigan Avenue, NW, Washington, DC, USA.
Objectives:
Gangliosides (GGs) are considered as diagnostic biomarkers and therapeutic targets and agents. The goal of this study was to develop a tandem mass spectrometry (MS/MS) method for the simultaneous measurement of both GM1 and GM2 gangliosides in human cerebrospinal fluid (CSF) samples in order to be able to determine their concentrations in patients with Tay-Sachs and Sandhoff disease and assess whether drugs or transplantation affect their concentrations.
Design And Methods:
An API-4000 tandem mass spectrometer equipped with TurboIonSpray source and Shimadzu HPLC system was employed to perform the analysis using isotope dilution with deuterium labeled internal standards. To a 1.5 mL conical plastic Eppendorf centrifuge tube, 40 microL of human CSF sample was added and mixed with 400 microL of internal standard solution for deproteinization. After centrifugation, 100 microL of supernatant was injected onto a C-18 column. After a 2.5 min wash, the switching valve was activated and the analytes were eluted from the column with a water/methanol gradient into the MS/MS system. Quantification by multiple reaction-monitoring (MRM) analysis was performed in the negative mode.
Results:
The within-day coefficients of variation were <3% for GM1 and <2% for GM2 and the between-day coefficients of variation were <5% for both GM1 and GM2 at all concentrations tested. Accuracy ranged between 98% and 102% for both analytes. Good linearity was also obtained within the concentration range of 10-200 ng/mL (6.5-129.3 nmol/L) for GM1 and 5-100 ng/mL (3.6-72.3 nmol/L) for GM2 (r> or =0.995).
Conclusions:
A new simple, accurate, and fast isotope dilution tandem mass spectrometry method was developed for the simultaneous quantification of GM1 and GM2 gangliosides in a small amount of human CSF. Concentrations were measured in "normal" CSF and in CSF from patients with Tay-Sachs disease.
More Related Videos
08:36Sublimation of DAN Matrix for the Detection and Visualization of Gangliosides in Rat Brain Tissue for MALDI Imaging Mass Spectrometry
Published on: March 23, 2017
08:53Quantitative and Qualitative Method for Sphingomyelin by LC-MS Using Two Stable Isotopically Labeled Sphingomyelin Species
Published on: May 7, 2018
Related Concept Videos
Tandem Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrometry: Complex Analysis